Image Dimension Measuring SystemIM series

IM-6120

Measurement Head: Wide-field model

[Discontinued model]

This model has been discontinued.

Recommended Replaceable Products
To contact us:

IM-6125 - Measurement Head: Wide-field model

Downloads

Catalogues

View Details
  • CE Marking

SPECIFICATIONS

Select Language

Model

IM-6120

Type

Measurement head

Image pickup device

1" 6.6 mega pixel CMOS

Display

10.4" LCD monitor (XGA: 1024 × 768)
External monitor connectable (clone output)

Light receiving lens

Double telecentric lens

Field of view

Wide field: ø100 × L200 mm, High precision: 25 × L125 mm

Minimum display unit

0.1 µm

Repetition accuracy

W/o stage movement

Wide field: ±1µm High precision: ±0.5µm

With stage movement

Wide field: ±2µm High precision: ±1.5µm

Measurement accuracy (±2σ)

W/o binding

Wide field: ±5µm*1 High precision: ±2µm*2

With binding

Wide field: ± (7 +0.02L)µm*3 High precision: ± (4 +0.02L)µm*4

External remote input

No-voltage input (with contacts, without contacts)

External output

OK/NG/FAIL

Relay outputRated load 24 VDC 0.5AON resistance 50 mΩ or less

LAN

RJ-45 (10BASE-T / 100BASE-TX / 1000BASE-T)

USB 2.0 series A

6 ports (front: 2, rear: 4)

Record

Hard disk drive

250 GB

Illumination system

Transparent

Telecentric transparent illumination (Green LED)

Ring

Four division ring illumination (White LED)

External

Externally connected illumination (coaxial)

X stage

Moving range

100 mm (electric)

Withstand load

2 kg

Z stage

Moving range

30 mm (electric)

Rating

Power voltage

100 to 240 VAC 50 / 60 Hz

Power consumption

310 mA or less

Environmental resistance

Operating ambient temperature

+10 to 35°C

Operating ambient humidity

20 to 80%RH (no condensation)

Weight

Approx 31 kg

*1 In the range of ø80 mm from the centre of the stage within the operating temperature range of +23°C ±1°C at the focused focal point position
*2 In the range of ø20 mm from the centre of the stage within the operating temperature range of +23°C ±1°C at the focused focal point position
*3 In the range of ø80 × 180 mm from the centre of the stage, within the operating temperature range of +23°C ±1°C at the focused focal point position, and with a load weighing 1 kg or less on the stage (L = amount of stage movement in mm units)
*4 In the range of ø20 × 120 mm from the centre of the stage, within the operating temperature range of +23°C ±1°C at the focused focal point position, and with a load weighing 1 kg or less on the stage (L = amount of stage movement in mm units)

Downloads:

For visitors who viewed this page

    Measurement Systems Library

    Latest / Most Downloaded

      • Trade Shows and Exhibitions
      • eNews Subscribe

      Measurement Systems