Laser scan micrometerLS-3000 series

LS-3033SO(8073)

LS-3033 Specially Ordered

[Discontinued model]

This model has been discontinued.

Recommended Replaceable Products
To contact us:

SPECIFICATIONS

Select Language

Model

LS-3033SO(8073)

Type

Fine measuring

Scanning head type

Separate

Controller

LS-3100W/LS-3100DW

Measuring range

0.08 to 30 mm *1

Minimum target width

0.08 mm

Transmitter/receiver distance

120 mm (Fixed) *2

Light source

Semiconductor laser, wavelength: 780 nm, output: 55μ, Class 1 laser product (JIS C6802)

Measurement accuracy

±2 μm max.*3

Repeatability
(number of averaging measurements: 512)

±0.3 μm

Minimum displayable unit

0.1μm

Laser scan rate

400 scans/s

Laser scan velocity

63 m/s

Laser scan range

Approx. 33 mm

Light source

Type

Invisible infrared semiconductor laser

Output

3.0 mW (FDA (CDRH) Part 1040.10), 55 μW (IEC 60825-1)

Wavelength

780 nm

Pulse duration

111 μs

Laser classFDA (CDRH) 21CFR Part 1040.10

Class I Laser Product*4

Laser class IEC 60825-1

Class 1 Laser Product*4

Display

Measurement display

Main display: 7-segment green LED, 8-digit sub display: 16-character LCD × 2 rows (backlight color: yellowish green)

Minimum display unit

0.1 μm

Display range

-999.9999 to 9999.9999

Position monitor

Green LED (7 levels)

Tolerance check output display

Green LED x 3 (HI/GO/LO)

Laser emission indicator

Green LED x 2 (HEAD 1/ HEAD 2)

Terminal block output

Control output

Analogue output voltage

±6 V, Output impedance: 0 Ω

Controloutput

SPST-NO Relay contact 250 VAC 2 A (resistive load)

Control input

Hold timing input

Non-voltage input (contact/solid state)

Laser remote input

Program switching input (program No. 1 to 5)

36-pin connector I/O

Hold timing input

TTL, level, negative logic

Auto-zero set/reset input

Program switching input (program No. 1 to 5)

Controloutput

7-level judgment output

Maximum NPN open-collector output: 40 mA (30 V or less)

3-step limits output

Interface

Type

RS-232C/GP-IB (optional)/ BCD (optional)

RS-232C

Measurement data output and control output (baud rate: selectable between 75 and 19,200 bps)

GP-IB

Measurement data output and control output*5

BCD

Measured data output (8 digits)*6

Rating

Power voltage

85 to 264 VAC, 50/60 Hz

Power consumption

40 VA max.

Environmental resistance

Enclosure rating

IP60

Operating ambient temperature

0 to 45°C

Operating ambient humidity

35 to 85%, No condensation

Weight

T: 0.5 kg
R: 0.4 kg*4

*1 When the outer diameter is measured.
*2 ‘T' indicates a transmitter, ‘R' indicates a receiver.
*3 When a rod 10 mm in diameter is measured in a 2 x 20 mm, 4 x 20 mm, or 10 x 20 mm measuring area.
*4 When measuring a rod 20 mm in diameter in a 20 x 40 mm measuring area.
*5 Optional accessories
*6 Supports options