          SPECIFICATIONS

          Select Language

          Model

          VK-X105

          Type

          Measurement unit

          Objective lens magnification

          Magnification on a 15” monitor 100x

          5x

          Magnification on a 15” monitor 200x

          10x

          Magnification on a 15” monitor 400x

          20x

          Magnification on a 15” monitor 1000x

          50x

          Observation/
          measuring range

          Horizontal (H): μm

          Magnification on a 15” monitor 100x

          2700

          Magnification on a 15” monitor 200x

          1350

          Magnification on a 15” monitor 400x

          675

          Magnification on a 15” monitor 1000x

          270

          Vertical (V): μm

          Magnification on a 15” monitor 100x

          2025

          Magnification on a 15” monitor 200x

          1012

          Magnification on a 15” monitor 400x

          506

          Magnification on a 15” monitor 1000x

          202

          Working distance: mm

          Magnification on a 15” monitor 100x

          22.5

          Magnification on a 15” monitor 200x

          16.5

          Magnification on a 15” monitor 400x

          3.1

          Magnification on a 15” monitor 1000x

          0.54

          Numerical Aperture (N.A.)

          Magnification on a 15” monitor 100x

          0.13

          Magnification on a 15” monitor 200x

          0.3

          Magnification on a 15” monitor 400x

          0.46

          Magnification on a 15” monitor 1000x

          0.8

          Optical zoom

          1x to 8x

          Total magnification

          100x to 8000x

          Optical system for observation/measurement

          Pinhole confocal optical system

          Height measurement

          Measuring range

          7 mm

          Z-resolution

          0.005 µm

          Repeatability σ

          0.02 µm*1

          Width measurement

          Display resolution

          0.01 µm

          Repeatability 3σ

          0.05 µm*2

          Frame memory

          Pixel count

          2048 x 1536, 1024 x 768, 1024 x 64

          For monochrome image

          16 bit

          For colour image

          8-bit for RGB each

          For height measurement

          21-bit

          Frame rate

          Surface scan

          4 Hz to 120 Hz

          Line scan

          7900 Hz

          Auto function

          AAG (Auto gain), Auto focus, Auto upper/lower limit setting, Double Scan brightness settings

          Measurement laser light source for measurements

          Wavelength

          Red semiconductor laser, 658 nm

          Maximum output power

          0.95 mW

          Laser Class

          Class 2 Laser Product (IEC 60825-1, FDA (CDRH) Part1040.10 *3)

          Laser light-receiving element

          PMT (Photoelectron Multiplier Tube)

          Light source for
          optical observation

          Lamp

          100 W halogen lamp

          Colour camera for
          optical observation

          Imaging element

          1/3" Colour CCD image sensor

          Recording resolution

          Super high resolution (3072×2304)

          Auto adjustment

          Gain, Shutter speed

          Data processing unit

          Dedicated PC, supplied by KEYENCE with the VK-X (OS: Windows 7 Professional Edition) *4

          Software

          Auto-focus software

          This is software that automatically focuses the image during observation.

          Pixel shift software

          This is software that increases the number of pixels recorded by shifting pixels during image recording.

          HDR function software

          This is software that obtains 16-bit data to optimize the image appearance.

          RPD function software

          This is an algorithm that detects the measurement result quickly and accurately.

          AAG function software

          This is software that automatically adjusts the laser light-receiving sensitivity within the setting range.

          Auto upper/lower limit setting software

          This is software that automatically recognizes the upper and lower limits of the target and sets the measurement range.

          Beginner mode function software

          This is software that can be used to easily perform measurements as flow charts.

          Comment function software

          This is software that displays comments on the observation screen.

          Measurement and scale display software

          This is software that displays a scale on the dimension measurement and observation screen.

          Rating

          Power voltage

          100 to 240 VAC ±10 %, 50/60 Hz

          Power consumption

          450 VA max.

          Weight

          Approx. 25 kg (Measuring head detached: approx. 8.5 kg)

          *1 When a standard height difference of 2 µm is measured with the 50x objective lens.
          *2 When the KEYENCE reference chart line width of 1 µm is measured with the 50 x objective lens (optical zoom 2 x) in line peak mode (image averaging: 8 times).
          *3 The laser classification for FDA (CDRH) is implemented based on IEC60825-1 in accordance with the requirements of Laser Notice No.50.
          *4 Windows 7 is a registered trademark of Microsoft, U.S.A.